《化学分析计量》 2019年第04期
DOI:10.3969/j.issn.1008-6145.2019.04.003
手持式X射线荧光光谱仪在牙科非贵金属无损检测中的应用
李稳,鄢祥,张殿云,林红
北京大学口腔医学院,口腔医院口腔材料研究室,口腔医疗器械检验中心,口腔数字医学北京市重点实验室,北京 100081 
Application of the portable X ray fluorescence spectrometer in nondestructive testing of dental non precious metals
李稳,鄢祥,张殿云,林红
Peking University and Hospital of Stomatology, Dental Medical Devices Testing Centre, Dental Material Research Center, Beijing Key Laboratory of Digital Stomatology, Beijing 100081, China 
摘要:采用手持式X射线荧光光谱(PXRF)法对两种口腔常用标准品和4种常用牙科非贵金属的主要元素进行无损检测,并且与原子吸收光谱(AAS)法进行比较。对YSBC18604-08钛铁标准品的主要元素Fe,Ti,Al,Mn和YSBC15343-2008不锈钢标准品的主要元素Fe,Cr,Mn以及4种已知元素含量的牙科钴铬支架(Co,Cr,Mo,W)、镍铬合金金属桥(Ni,Cr,Mo)、齿科钴铬烤瓷合金(Co,Cr,Mo,W)、镍基铸造合金(Ni,Cr,Cu,Sn)进行分析检测。对含量较高(>20%)的元素,其测定结果的相对标准偏差均小于0.8%(n=9),含量小于10%的元素其测定结果的相对标准偏差均小于4%(n=9)。不锈钢标准品YSBC15343-2008和钛铁标准品YSBC18604-08各元素测定结果与标称值基本一致,相对偏差在0.025%~2.54%范围内。PXRF法和AAS法对原材料中主要元素的测定结果基本一致,均与标称含量相符合。PXRF法对成品及原材料的分析不需要破坏试样,分析速度快,准确度高,重现性好,能同时检测多种元素,满足日常分析要求。 
Abstract: The portable X-ray fluorescence analysis (PXRF) method was used for the nondestructive detecting of the main elemental contents in two kinds of standard structural specimen and four kinds of dental used non precious metallic materials. The elemental contents results detected by PXRF method and atomic absorption spectrometry (AAS) were compared. Fe, Ti, Al, Mn in the YSBC18604-08 titanium-iron standard sample, Fe, Cr, Mn in the YSBC15343-2008 stainless steel standard sample, Co, Cr, Mo, W in dental stent, Ni, Cr, Mo in mental bridge, Co, Cr, Mo, W in dental cobalt chromium ceramic alloy and Ni, Cr, Cu, Sn in the nickel based cast alloy were tested by PXRF method. As the elemental content was more than 20%, the relative standard deviation of detection results was less than 0.8%(n=9), while was less than 4%(n=9) as the elemental content was less than 10%. The detection results of target elements in YSBC15343-2008 stainless steel and YSBC18604-08 titanium-iron standard samples were close to the nominal values, the relative deviation in the rang of 0.025%-2.54%. The results of major elements in raw materials detected by PXRF and AAS method showed no difference, and all being close to the nominal values. This method can satisfy the ordinary analysis, it is nondestructive test, and cost less, besides, more than one kind of elements can be detected simultaneously. 
关键词:手持式X射线荧光光谱法;牙科非贵金属;无损检测;原子吸收光谱法 
Keywords:portable X-ray fluorescence analysis method; dental non precious metal; nondestructive testing; atomic absorption spectrometry 
本文引用格式:
李稳,鄢祥,张殿云. 手持式X射线荧光光谱仪在牙科非贵金属无损检测中的应用[J].化学分析计量,2019,28(4):13-17.
LI W,YAN X,ZHANG D Y,et al. Application of the portable X ray fluorescence spectrometer in nondestructive testing of dental non precious metals[J]. Chemical analysis and meterage,2019,28(4):13-17. 

全文:
相似文章:
  快速消解-石墨炉原子吸收法测定鸡蛋中铅、镉、铬、铜含量
2024年第02期
  乙醇和氯化铵提取-原子吸收光谱法测定石灰性土壤中的交换性钾、钠、钙、镁
2024年第01期
  微波辅助提取原子吸收光谱法快速检测食品中铅、镉、锰、锌
2023年第10期
  氧化还原体系溶液进样对原子吸收火焰的增温作用
2023年第05期
  石墨炉原子吸收法测定虫草花中铅含量的不确定度评定
2023年第04期
  火焰原子吸收光谱法测定预脱碳液中的铬
2022年第08期
  火焰原子吸收光谱法测定镁及镁合金中镉含量不确定度评定
2022年第03期
  钯含量分析方法研究进展
2021年第12期
  乳粉中钙成分分析标准物质研制
2021年第05期
  催化热解-塞曼原子吸收光谱法测定茶叶中的汞
2021年第04期
《化学分析计量》杂志社 版权所有,未经授权禁止复制或者建立镜像
地址:山东省济南市天桥区田庄东路3号 邮编:250031
电话:0531-85878132 85878224 85878278
鲁ICP备05018524号-6