电感耦合等离子体原子发射光谱法测定压水反应堆硼酸介质中的铬、镍、钴
郑德超1,韦荣伟1,刘晓军1,罗坤1,陈思羽2,许发伦1,陈乾茂3
1.广西防城港核电有限公司,广西防城港 538000; 2.中核深圳凯利核能服务有限公司防城港分公司,广西防城港 538000; 3.珀金埃尔默企业管理(上海)有限公司,上海 201203 
Determination of Cr, Ni and Co in boric acid medium of pressurized water reactor by inductively coupled plasma atomic emission spectrometry
ZHENG Dechao1, WEI Rongwei1, LIU Xiaojun1, LUO Kun1, CHEN Siyu2, XU Falun1, CHEN Qianmao3
1. Guangxi Fangchenggang Nuclear Power Co., Ltd., Fangchenggang 538000, China; 2. Fangchenggang Branch, CNNC Shenzhen Kaili Nuclear Power Service Co., Ltd., Fangchenggang 538000, China; 3. PerkinElmer Enterprise Management (Shanghai) Co., Ltd., Shanghai 201203, China 
摘要:建立电感耦合等离子体原子发射光谱法测定压水反应堆硼酸介质中铬、镍、钴含量的分析方法。分析谱线为Cr 206.158 nm,Ni 221.648 nm,Co 231.160 nm。对样品作酸化处理,使用线性加权校正,采用基本匹配和多谱线拟合技术联用消除光谱干扰。对于硼基体浓度为0~2 500 mg/kg的样品,铬、钴的线性范围为10~200 μg/kg,镍的线性范围为40~800 μg/kg,线性相关系数均大于0.999,铬、镍、钴的检出限分别为0.81,0.67,0.67 μg/kg。测定结果的相对误差不大于7.7%,测定结果的相对标准偏差不大于3.1%(n=6),加标回收率为91.5%~108.5%。该方法检测快速,测定结果准确,光谱干扰校正模型稳定,能满足实际生产需要。 
Abstract:The method for determining Cr, Ni and Co in the boric acid medium of pressurized water reactor by inductively coupled plasma atomic emission spectrometry was established. The spectral lines were Cr 206.158 nm, Ni 221.648 nm and Co 231.160 nm. The samples were acidified, linear weighted correction was used, basic matching and multispectral line fitting technology were used to eliminate spectral interference. For the sample with boron matrix concentration of 0-2 500 mg/kg, the linear correlation coefficients were more than 0.999, the linearity range of Cr and Co was 10-200 μg/kg, and that of Ni was 40-800 μg/kg, the detection limit of Cr, Ni and Co was 0.81, 0.67, 0.67 μg/kg, respectively. The relative errors of measurement results were not more than 7.7%. The relative standard deviations of determination results were not more than 3.1%(n=6). The standard recovery of each element was 91.5%-108.5%. The method is fast and accurate, and the spectral interference correction model was stable. The method can satisfy the actual production requirements. 
关键词:活化腐蚀产物;电感耦合等离子体原子发射光谱法;多谱线拟合技术;铬;镍;钴 
Keywords:activated corrosion product; inductively coupled plasma atomic emission spectrometry; multispectral line fitting technology; Cr; Ni; Co 
本文引用格式:
郑德超,韦荣伟,刘晓军,等.电感耦合等离子体原子发射光谱法测定压水反应堆硼酸介质中的铬、镍、钴[J].化学分析计量,2019,28(5):76-80. ZHENG D C,WEI R W,LIU X J,et al. Determination of Cr,Ni and Co in boric acid medium of pressurized water reactor by inductively coupled plasma atomic emission spectrometry[J]. Chemical analysis and meterage,2019,28(5):76-80. 

全文:
相似文章:
  电感耦合等离子体原子发射光谱法同时测定不锈钢中的11种元素
2019年第06期
  电感耦合等离子体质谱法同时测定药用辅料山梨醇中的铅和镍
2019年第06期
  微波消解-电感耦合等离子体质谱法同时测定垃圾渗滤液中8种生物毒性元素
2019年第06期
  基于电感耦合等离子体高纯镁基氧化物中共存元素光谱特征
2019年第06期
  电感耦合等离子体发射光谱法测定不锈钢中的铬、镍、锰、铜、钛、铝
2019年第05期
  电感耦合等离子体原子发射光谱法测定牙科修复体产品中的铍、镉
2019年第05期
  电感耦合等离子体发射光谱法测定化妆品中可溶性钡与镍
2019年第05期
  高效液相色谱-电感耦合等离子体质谱法测定印刷油墨中可迁移Cr(Ⅲ)和Cr(Ⅵ)
2019年第05期
  ICP-AES法测定镍铬铝钇硅合金中的铝、钇、硅
2019年第04期
  电感耦合等离子体原子发射光谱法测定合金钢中的铼
2019年第04期
《化学分析计量》杂志社 版权所有,未经授权禁止复制或者建立镜像
地址:山东省济南市天桥区田庄东路3号 邮编:250031
电话:0531-85878132 85878148 85878224 85878278 传真:0531-85947355
鲁ICP备05018524号-6