收藏我们   ENGLISH  
论文查询
范围:  
提 交
《化学分析计量》 2022年第11期
DOI:10.3969/j.issn.1008-6145.2022.11.019
高效液相色谱法测定婴幼儿用奶瓶中己内酰胺迁移量的不确定度评定
杨芳,梁婷婷
淮安市产品质量监督综合检验中心,江苏淮安 223001 
Uncertainty evaluation of determination of caprolactam migration in baby bottle by high performance liquid chromatography
YANG Fang, LIANG Tingting
Huaian Product Quality Supervision and Inspection Center, Huaian 223001, China 
摘要:采用高效液相色谱法测定婴幼儿用奶瓶中己内酰胺迁移量,并对其测定的结果进行不确定度的评估。参照JJF 1059.1—2012《测量不确定度评定与表示》和CNAS—GL 006:2019《化学分析中不确定度的评估指南》中的规定与要求,分析己内酰胺在测定过程中的各种不确定度来源。水基食品模拟物处理的样品中己内酰胺的迁移量测定结果为12.213 mg/L,其扩展不确定度为1.29 mg/L,结果表示为(12.21±1.29) mg/L, k=2;酸性食品模拟物处理的样品中己内酰胺的迁移量测定结果为15.608 mg/L,扩展不确定度为1.48 mg/L,结果表示为(15.61±1.48) mg/L, k=2;酒精类食品模拟物处理样品中己内酰胺所产生的迁移量测定结果为18.299 mg/L,扩展不确定度为1.84 mg/L,结果表示为(18.30±1.84) mg/L,k=2;油基食品模拟物处理样品中,己内酰胺的迁移量测定结果为10.232 mg/kg,扩展不确定度为0.99 mg/kg,结果表示为(10.23±0.99) mg/kg, k=2。通过评估发现,不确定度主要的影响因素是标准溶液的配制过程、仪器分析和样品重复性测量。 
Abstract:The uncertainty of the determination of caprolactam migration in baby bottle by high performance liquid chromatography(HPLC) was evaluated. According to the Provisions and Requirements in Evaluation and Expression of Measurement Uncertainty (JJF 1059.1-2012) and Guidance on Quantifying Uncertainty in Chemical Analysis (CNAS-GL 006: 2019), the sources of various uncertainties in the determination process of caprolactam were analyzed. On water simulated fluid, the migration of caprolactam was 12.213 mg/L with the expanded uncertainty of 1.29 mg/L, the result was expressed as (12.21±1.29) mg/L, k=2. On acid simulated fluid, the migration of caprolactam was 15.608 mg/L with the expanded uncertainty of 1.48 mg/L, the result was expressed as (15.61±1.48) mg/L, k=2. On ethanol simulated fluid, the migration of caprolactam was 18.299 mg/L with the expanded uncertainty of 1.84 mg/L, the result was expressed as (18.30±1.84) mg/L, k=2. On oil simulated fluid, the migration of caprolactam was 10.232 mg/kg with the expanded uncertainty of 0.99 mg/kg, the result was expressed as (10.23±0.99) mg/kg, k=2. The main factors affecting the uncertainty are sample pretreatment, preparation of standard solution and linear fitting process of standard solution 
关键词:高效液相色谱法;婴幼儿用奶瓶;己内酰胺迁移量;不确定度 
Keywords:high performance liquid chromatography; baby bottle; caprolactam migration; uncertainty 
本文引用格式:
杨芳,梁婷婷.高效液相色谱法测定婴幼儿用奶瓶中己内酰胺迁移量的不确定度评定[J].化学分析计量,2022,31(11):90.
YANG Fang,LIANG Tingting. Uncertainty evaluation of determination of caprolactam migration in baby bottle by high performance liquidchromatography[J]. Chemical Analysis and Meterage,2022,31(11):90. 

全文:
相似文章:
  Top-down控制图法评定石墨炉原子吸收光谱法测定血中铅含量的不确定度
2024年第02期
  气相色谱法测定酱油中对羟基苯甲酸甲酯的不确定度评定
2024年第02期
  Top-down法评定EDTA络合滴定法测定海洋沉积物中碳酸钙含量的不确定度
2024年第01期
  超高效液相色谱-串联质谱法检测蘑菇中6种蘑菇毒素的不确定度评定
2024年第01期
  电喷雾检测器高效液相色谱法测定复方北豆根氨酚那敏片中灰毡毛忍冬皂苷乙和川续断皂苷乙
2024年第01期
  高效液相色谱法同时测定明胶空心胶囊中苯甲酸和山梨酸
2024年第01期
  高效液相色谱法测定化妆品原料中N-乙酰神经氨酸
2024年第01期
  高效液相色谱法测定诃子中诃黎勒酸和诃子酸
2024年第01期
  多功能针式过滤器净化超高效液相色谱法测定大米和花生中玉米赤霉烯酮
2024年第01期
  基于WPSVBA编程的Top-Down法评定化学分析测量不确定度
2023年第11期
《化学分析计量》杂志社 版权所有,未经授权禁止复制或者建立镜像
地址:山东省济南市天桥区田庄东路3号 邮编:250031
电话:0531-85878132 85878224 85878278
鲁ICP备05018524号-6